Continuous automatic measurement system for solid/powder samples for 20 samples.
Equipped with automatic reagent exchange function.
Create a system by combining it with a KF Titrator. A series of operations such as sample container transfer, heating, and measurement are performed automatically.
Moisture Vaporization Method: Ventilation heating vaporization method
Sample Container: Hard glass bottle, aluminum foil packing, PTFE lid
Number of Sample Containers: 20 samples
Sample Amount (Volume): Approximately 10 mL or less
Carrier Gas:
Reagent Exchange: Automatic exchange by setting the number of measurements
Carrier Gas Desiccant: Molecular sieves
Heater / Temperature Control: Band heater with PID control method
Temperature Range: Room temperature to 300°C (up to 3 temperature levels can be set)
End Position Setting: Operation stops when the stop pin reaches the end position or when no sample container is detected
Post-Measurement Procedures: End buzzer notification
Power Supply: AC 100–240 V, 50/60 Hz, approximately 150 VA
*A power cable is not included with this product.
Dimensions / Weight: